Accurate elemental analysis starts long before the sample reaches the instrument. For ICP-OES, ICP-MS, and AAS, solid samples such as soil, food, pharmaceuticals, polymers, and biological tissue must first be converted into a clear liquid. Microwave digestion is the most widely used way to do this, because it is fast, reproducible, and works in a closed system. This guide explains how it works and how to get reliable results. You can browse our full range of microwave digestion products at any time.
In microwave digestion, a sample is sealed in a pressure-resistant vessel together with an acid or acid mixture. Microwave energy heats the contents rapidly, and the sealed vessel allows temperatures well above the normal boiling point of the acids. Under these conditions, organic matrices break down and metals dissolve into solution within minutes to a couple of hours, depending on the sample.
Systems such as the CEM MARS 6 are designed for exactly this kind of closed-vessel work in analytical and quality control labs.
| Factor | Microwave (closed vessel) | Open hot-plate digestion |
|---|---|---|
| Speed | Typically under 1 to 2 hours | Often several hours |
| Volatile elements | Retained in the sealed vessel | Risk of loss (e.g., Hg, As, Se) |
| Contamination | Lower, with less acid and less exposure to air | Higher |
| Acid consumption | Low | High |
| Reproducibility | Excellent, with controlled temperature and pressure | Operator-dependent |
| Safety | Fume exposure is reduced | More exposure to acid fumes |
Always match the acid mixture to your sample type and your published or validated method. High-purity reagents and trace-metal-grade chemicals from brands such as Merck and Supelco are commonly used for low-level work.
Many labs follow established methods, for example:
Always check the current version of the method you are following and confirm the required temperature and hold times.
Plastic tubes, flasks, and other laboratory consumables used for dilution and storage should also be low in trace metals; otherwise, they can raise your blanks.
Incomplete digestion (cloudy solution or residue)
High blanks
Vessel venting or leaks
Poor reproducibility
Our CEM system support section lists spare parts and accessories to keep your digestion system running safely. For manuals and technical documents, visit our help documents page.
Clean digestion only pays off if the ICP-MS interface is also in good condition. Dirty or worn cones and sample-introduction parts cause signal drift and poor detection limits, especially after running high-matrix digests. Consumables such as the high-performance nickel skimmer cone for Thermo Fisher instruments are part of regular ICP-MS maintenance. You can find more parts from Thermo Fisher and other brands on our manufacturer pages.
Microwave digestion involves strong acids and high pressure. Always wear suitable protective equipment, work in a ventilated area, follow the manufacturer's maximum load and program limits, and let vessels cool before opening. HF requires specific training and first-aid preparation.
Ekelabshop supplies spare parts and consumables for CEM microwave digestion systems, including the MARS 6 platform, along with a wide range of laboratory supplies from leading manufacturers. If you need vessels, liners, caps, seals, sensors, or other replacement parts, send us the part number or the model of your system, and we will help you find the right match.