A Thermo 49i pump rebuild means replacing the pump’s diaphragm and valve plate. It restores the analyzer’s vacuum and flow without buying a whole new pump. For most iSeries analyzers, the job takes one kit, a few hand tools, and careful attention to the order of the parts. This guide explains how to confirm your pump type, pick the right kit, and carry out the work.
The internal pump does three jobs at once. It pulls the sample through the analyzer, draws ambient air through the ozone generator, and keeps the vacuum the measurement needs. The internal pump serves three essential duties: it draws the sample through the analyzer, draws ambient air through the ozone generator, and maintains the vacuum for the reaction.
Because every one of those functions depends on one small diaphragm pump, a worn diaphragm or tired valve plate can affect the whole measurement, not just the airflow.
A rebuild makes sense when the pump motor still runs smoothly, but the pumping action has weakened. A diaphragm and its valves are wear parts, and they are what the rebuild kit replaces. Typical signs are low flow readings, pressure values drifting out of their normal range, or a pump that runs but struggles to pull the expected vacuum. Check these in the analyzer’s diagnostics before opening anything, because capillaries, filters, and leaks can produce similar symptoms.
Thermo also treats the pump rebuild as routine maintenance. Its guidance for shutting down a 49i for the off-season lists changing the filter, rebuilding the pump, cleaning the cells, checking the lamp, and inspecting the scrubber. If you want to plan the other wear parts around the same service visit, our Thermo iSeries spare parts replacement guide covers when to replace the pump, diaphragm, pressure sensor, and filter tape.
If the motor is noisy, the bearings feel rough, or the pump housing is damaged, a rebuild kit will not fix the problem. In that case, replacing the whole pump is the better choice.
A rebuild kit is the cheaper route when only the diaphragm and valves have aged. A complete pump assembly is the better route when the motor or bearings are the problem, or when you want to swap the pump quickly and rebuild the old one later on the bench. For a swap, the 111553-01 Thermo 49i pump assembly is the complete replacement pump. For a rebuild, you need the Thermo 108002-00 pump rebuild kit.
| Situation | Better choice |
|---|---|
| Weak flow, motor and bearings fine | Rebuild kit |
| Rough bearings or noisy motor | Complete pump |
| Analyzer must be back online quickly | Complete pump, rebuild the old one afterward |
| Keeping a spare-parts stock | Kit, plus one spare pump for critical sites |
Before ordering, read the part number on your pump and match it to the correct kit. Thermo’s iSeries pump bulletin states that rebuild kit 108002-00 is shared by pump part numbers 111553-01, 111553-00 and 101426-00, and that it includes replacement shims. The same bulletin lists the 49i and 49i-PS, along with the 43i, 48i, and 146i, as models that use these pumps.
The shims are where mistakes happen. In the same bulletin, the 111553 pumps are low-vibration, shim-less designs, so the shims in the kit should be discarded. The 101426-00 pump comes in two forms: the “old style” is shim-less and uses a different kit, 8606, while the “new technology” version uses shims. The old shims you remove must go back in the same quantity, so keep them in order.
| Pump part number | Rebuild kit | Shims |
|---|---|---|
| 111553-01 / 111553-00 | 108002-00 | None; discard the kit shims |
| 101426-00 “new technology” | 108002-00 | Required; reuse as removed |
| 101426-00 “old style” | 8606 | None |
Note that this bulletin is dated December 2011, so confirm current part numbers with your supplier before ordering. The 42i family uses a different pump and kit (107751-00), so do not assume the same kit covers every analyzer in the series.
You need very little: a #1 Phillips screwdriver, a 2.5 mm Allen key, a marker or pencil, and the rebuild kit. Thermo’s procedure also calls for Teflon tape when refitting fittings.
Before you start:
The steps below follow Thermo’s published procedure for the 111553 and 101426 pumps. Always read the instruction sheet supplied with your kit, since it governs your specific pump.
Refit the compressor housing cover and gasket, but do not over-tighten the cover screws. Remove all old Teflon tape from the fittings and apply fresh tape. Thermo advises two layers, kept on the threads only, because stray tape can tear off and lodge in the valves. Reconnect the tubing from your sketch, and avoid over-tightening metal fittings.
After reinstalling the pump, power the analyzer and check flow and pressure readings in the diagnostics menu. Compare them with the values you recorded before the rebuild, or with your analyzer manual’s normal ranges. Listen for unusual noise, and check the fittings for leaks. If flow is still low, look at the capillaries and the sample filter next, since the pump is no longer the only suspect.
Run your normal calibration or verification routine before returning the instrument to service. A pump change can affect the sample path, so treat it as a maintenance event that needs documenting.
Yes, for the common iSeries pumps. Thermo’s bulletin lists it for pumps 111553-01, 111553-00 and 101426-00 used in the 49i, 49i-PS, 43i, 48i and 146i. Check your pump label first, because the “old style” 101426-00 uses kit 8606.
It depends on the pump. The 111553 pumps do not use shims, while the “new technology” 101426-00 does. Reuse shims only where the pump originally had them.
A #1 Phillips screwdriver, a 2.5 mm Allen key, a marker, and the kit. Teflon tape is needed for the fittings.
Thermo’s bulletin does not give a time, and it varies with experience. Allow time for careful disassembly, drawing your tube sketch, and testing afterward.
Rebuild when the motor and bearings are healthy and only the diaphragm and valves are worn. Replace the pump if the bearings or motor are damaged.