The Thermo 42i measures nitric oxide (NO) and total nitrogen oxides (NOx) by chemiluminescence. Ozone reacts with NO inside a reaction chamber; the reaction gives off a faint light, and a photomultiplier tube (PMT) turns that light into a signal. The parts that wear out most often are the capillaries, O-rings, pump diaphragm, ozone generator, and molybdenum converter.
If your 42i has started to read low, drift, or throw flow and temperature alarms, the cause is almost always one of those wear parts. Knowing how the analyzer works makes it much easier to work out which one. This guide explains the measurement principle in plain language, lists the main components, and shows what usually fails first.
Always check your Thermo Scientific manual for exact flow rates, temperatures, and service intervals for your firmware and configuration. The figures below are general guidance.
The 42i uses a method called chemiluminescence. It sounds technical, but the idea is simple.
Nitric oxide reacts with ozone (O₃) and forms nitrogen dioxide (NO₂) in an excited state. As those excited molecules settle back to normal, they release a small amount of light. The more NO in the sample, the more light you get. A PMT sits next to the reaction chamber and measures that light. The analyzer converts the signal into a concentration.
Here is the sequence step by step:
The light signal only comes from NO. To see NO₂, the analyzer needs to turn it into NO first. It does this with a heated molybdenum converter.
In NO mode, the sample goes straight to the reaction chamber. In NOx mode, the sample passes through the converter first, which changes NO₂ to NO, so the reading includes both. The analyzer switches between these paths with a solenoid valve. NO₂ is then calculated as the difference between the NOx and NO readings.
Component | What it does | Typical problem |
|---|---|---|
Sample capillary | Sets sample flow | Partial blockage, low flow |
Ozone capillary | Sets ozone flow | Blockage, weak reaction |
Ozone generator | Makes ozone from dry air | Output drops with age |
Reaction chamber | Where NO and O₃ react | Contamination, leaking seals |
PMT and cooler | Detects light, kept cold | Noise, temperature alarms |
Molybdenum converter | Turns NO₂ into NO | Poor conversion efficiency |
Pump | Pulls gas through the system | Weak vacuum, noisy operation |
O-rings and fittings | Seal the flow path | Leaks, unstable readings |
Capillaries are tiny, precisely sized orifices, so dust or moisture can partly block them. The sample capillary (the larger, 15 mil purple one) and the ozone capillary (the smaller, 8 mil red one) are different sizes, so never mix them up. A partly blocked capillary lowers flow, and lower flow means a lower reading. See the 4127 sample capillary and the 4119 ozone capillary.
Viton O-rings harden and flatten over time, especially on the fittings that get opened during service. A small leak lets room air into the flow path, and the readings become unstable or low. Replacing the Viton O-rings on the Cajon fitting whenever you open the fitting is cheap insurance.
The pump works constantly. Eventually the diaphragm stiffens or cracks, and the pump can no longer pull the vacuum the analyzer needs. You may notice a flow alarm, a louder pump, or a slow response. You can read about the rebuild process in our Thermo 49i pump rebuild guide, and the same principles apply. Or you can swap the whole 101011-00 pump assembly.
Ozone output slowly falls with age. When it drops far enough, not all the NO reacts, and sensitivity suffers. Because the generator also has a drying and flow path around it, a low ozone flow alarm can point to a blocked capillary or a leak instead. Check those first. If they are fine, an ozone generator assembly is the next suspect.
The PMT itself lasts a long time, but its cooler works hard. If the cooler fails, the PMT warms up, background noise rises, and you may see a temperature alarm. The 9995 PMT cooler is the usual fix.
The converter slowly loses efficiency as the molybdenum is used up. If your NO reading is normal but NO₂ looks too low, converter efficiency is the first thing to test.
Use this as a quick starting point before you open the instrument.
Symptom | Likely cause | First thing to check |
|---|---|---|
Low sensitivity on NO and NOx | Blocked capillary, weak ozone | Flows, then capillaries |
Low NO₂ only | Converter efficiency | Converter age and temperature |
Noisy or drifting baseline | Leak or warm PMT | O-rings, fittings, cooler |
Flow alarm | Pump or blocked capillary | Pump vacuum, capillaries |
PMT temperature alarm | Cooler fault | Cooler and its fan |
Slow response | Weak pump, dirty lines | Pump, tubing, filter |
Start with flows and leaks. They are the cheapest to check and cause the most faults.
There is no single schedule that fits every site. Dusty, humid, or high-sulfur sources wear parts faster than clean ambient air monitoring. As a practical routine:
Keep a small stock of capillaries, O-rings, and a pump diaphragm kit. These are the cheap parts that cause the most downtime.
It measures NO, NO₂, and NOx in air. NO₂ is calculated from the difference between NOx and NO.
It is light produced by a chemical reaction. In the 42i, NO reacts with ozone and emits light that the PMT measures.
Only NO makes light with ozone. The converter changes NO₂ into NO so the analyzer can count total NOx.
Replace them when flow checks show they are blocked or out of range. Sites with dirty sample gas need to check them more often.
The most common causes are a blocked capillary, an air leak, a weak pump, or low ozone output.
If you have found the likely fault, you can browse our full range of Thermo iSeries and CEMS spare parts, or read our guide to when to replace iSeries pumps, sensors, and filters. Send us the part number or instrument model through the quote form and our team will help you match the right part.