If your Thermo iSeries analyzer is showing a low flow alarm or unstable readings, a worn or blocked capillary is one of the first parts to check. It is small and inexpensive, but it controls how much sample gas reaches the reaction chamber. Choosing the wrong one can throw off the whole measurement.
This guide explains what a Thermo iSeries capillary does, how the color codes and mil sizes work, when to replace one, and how to order the correct part.
In iSeries gas analyzers such as the 42i, 43i, 48i, 49i, and 450i, the capillary is a small calibrated orifice in the sample flow path. The internal pump pulls sample gas through it, and its bore size sets the flow rate. A narrower opening gives lower flow, and a wider one gives higher flow. To see how the flow path fits into the whole instrument, read our guide on how the Thermo 42i NOx analyzer works and which parts wear out.
This matters because the analyzer's readings depend on a stable, known sample flow. If the capillary is partly blocked, worn, or the wrong size, the flow changes. That can lead to flow alarms, drifting readings, slow response, or calibration that will not hold.
Thermo marks capillaries by size, and the color helps you identify the bore at a glance. Size is given in mils, where one mil is one thousandth of an inch. A larger mil number means a larger opening.
Part number | Size | Color | Listed application |
|---|---|---|---|
8 mil | Red | 42i NOx analyzer | |
15 mil | Purple | 42i NOx analyzer | |
20 mil | Blue | iSeries analyzers (confirm model) |
The color is a quick visual check, but do not order by color alone. Always match the part number to your analyzer model and instrument version.
Different iSeries models, and sometimes different versions of the same model, use different capillary sizes. Follow these steps before you order.
1. Identify your analyzer model. Note whether you have a 42i, 43i, 48i, 49i, or 450i. A 42i NOx analyzer and a 450i SO2/H2S analyzer do not necessarily use the same capillary.
2. Check the manual's parts list. The instrument manual lists the capillary part number for each flow path. This is the most reliable source.
3. Read the old part. If you can safely remove the existing capillary, check its color and any markings. A photo of the old part helps when you contact a supplier.
4. Note the serial number. Instrument versions change over time. The serial number lets a supplier verify that a replacement matches your analyzer.
If the manual and the old part disagree, ask your supplier to confirm before ordering. A wrong capillary is a cheap mistake, but the downtime it causes is not.
A capillary does not fail dramatically. It usually degrades slowly, so watch for these signs:
A blocked capillary is not the only cause of these symptoms. Pump wear, leaks, a clogged particulate filter, and a faulty pressure sensor can look similar. Check the filter and the pump first, then the capillary. Our Thermo iSeries spare parts guide explains when to replace the pump, diaphragm, pressure sensor, and filter tape, and our Thermo 49i pump rebuild guide walks through the diaphragm and valve replacement step by step.
Because the bore is so small, cleaning is unreliable. A partial clean can leave the opening slightly different from its original size, and the flow will no longer match the calibration. For this reason, most technicians replace the capillary rather than clean it. Follow the procedure in your instrument manual, and replace the part if there is any doubt.
Always follow your instrument manual for the full procedure. These points apply in most cases:
Many flow problems blamed on the capillary are really sealing problems. The O-rings in the capillary holder compress with every service and age with heat and exposure to sample gas. If you are already replacing the capillary, replace the O-rings at the same time. It adds little cost and removes a common source of leaks. You can request the 4800-PK O-ring for Thermo iSeries together with your capillary order.
When you request a quote, send:
With this information, the supplier can check that the replacement suits your instrument version. Ekelabshop supplies Thermo iSeries spare parts with worldwide shipping. You can browse our CEMS and air analyzer spare parts or contact us for pricing and lead time.
It is a small calibrated orifice that controls how much sample gas flows through the analyzer. Its bore size sets the flow rate.
The color identifies the bore size. For example, red marks the 8 mil 4119, purple the 15 mil 4127, and blue the 20 mil 4126. Always confirm the part number for your model.
There is no fixed interval. Replace it when you see flow alarms or unstable readings that trace back to the capillary, and follow your preventive maintenance schedule.
No. The capillary size is matched to the analyzer's design and calibration. Changing it will make the flow and the readings incorrect.