Use stainless steel tubing for UHPLC and high-pressure methods (above roughly 5,000 psi / 350 bar) or when your mobile phase contains THF, DMSO, or chlorinated solvents. Use PEEK tubing for metal-sensitive samples, high-chloride buffers, and moderate-pressure routine HPLC where easy finger-tight installation matters.
HPLC tubing carries the mobile phase and sample between the pump, autosampler, column, and detector. Fittings (nuts and ferrules) connect that tubing to each port and create a leak-free seal. Together, they form the flow path, and their material, inner diameter (ID), and connection quality directly affect pressure, peak shape, and reproducibility.
PEEK (polyether ether ketone) is a chemically resistant, metal-free polymer. Stainless steel is a rigid, high-pressure metal. The main differences are pressure rating, chemical compatibility, metal interaction and installation.
Factor | PEEK | Stainless Steel |
|---|---|---|
Typical pressure range | Up to ~5,000 psi (350 bar), depends on ID | 10,000 psi and above (UHPLC) |
Metal-free flow path | Yes | No |
High-chloride / high-salt buffers | Excellent | Limited (corrosion risk) |
THF, DMSO, chlorinated solvents | Poor (swelling or degradation) | Good |
Installation | Finger-tight, cut to length in the lab | Tools needed, ferrule swaged on |
Relative cost | Lower | Higher |
Best for | Biomolecules, chelating analytes, routine HPLC | UHPLC, aggressive organic solvents |
Choose PEEK when:
Choose stainless steel when:
Narrower tubing reduces extra-column band broadening but raises backpressure and clogs more easily. Internal volume per 10 cm of tubing:
Inner diameter | Volume per 10 cm | Typical use |
|---|---|---|
0.13 mm (0.005") | ~1.3 µL | UHPLC, small-volume columns |
0.18 mm (0.007") | ~2.5 µL | Modern HPLC, all-round choice |
0.25 mm (0.010") | ~5.1 µL | Conventional 4.6 mm ID column setups |
Rule of thumb: use the narrowest ID your system pressure allows, and keep the tubing between column and detector as short as practical.
Replace them when you see:
Your situation | Best choice |
|---|---|
UHPLC above ~6,000 psi | Stainless steel (or titanium / MP35N for corrosive mobile phases) |
Metal-sensitive or biological samples | PEEK or other metal-free tubing |
High-chloride buffers or extreme pH | PEEK |
THF, DMSO or chlorinated solvents | Stainless steel |
Routine lab with frequent column changes | PEEK finger-tight fittings |
Yes, for moderate-pressure methods. PEEK is metal-free, corrosion-resistant, and easy to install, but it is not suitable for UHPLC pressures or solvents such as THF, DMSO, and methylene chloride.
Standard PEEK tubing is typically rated to about 5,000 psi (350 bar), varying with inner diameter. Always check the manufacturer’s rating.
Generally no. THF can swell and weaken PEEK. Use stainless steel for THF-containing mobile phases.
Some analytes, such as phosphorylated and chelating compounds, adsorb onto metal surfaces in the flow path, which leads to tailing, low recovery, or poor reproducibility. A metal-free flow path often fixes this.